Techno-path Manufacturing recalls Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
What should I do?
Check whether you have this device (see the product and lot details below). If you do, stop using it and contact the company or your healthcare provider for next steps.
What's the hazard?
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
What's affected
Techno-Path Multichem AE Unassayed Level 2 REF: AE622X